Sulfimax GX Lab

Sulfimax GX Lab 型號:Sulfimax GX Lab

H2S analyzer for determination of hydrogen sulphide in liquids and gases - laboratory version, automatable with autosampler for liquids

  • Determination of hydrogen sulphide in liquids and gases

    The Sulfimax GX Lab fulfils the requirements of the German standard DIN 38405-27: Determination of sulphide by gas extraction method (D 27)

    The Sulfimax GX Lab determines hydrogen sulphide (H2S) and volatile sulphides in liquids and gases.

    The sample can be dosed directly without pretreatment - manually by syringe or in the automatic version by autosampler.

    By effective gas extraction the H2S is completely expelled from the sample. Interferences due to the sample matrix practically do not occur.

    The released H2S gas is conducted to the highly sensitive sensor, which detects H2S in the range of 0.01 to 10,000 ppm. A typical measurement takes 5 min, depending on the sample composition.

    If the Sulfimax GX Lab is extended with the optionally available H2S Headspace Module, solid and pasty samples can also be measured.

    Features

    • Complete separation of H2S from the sample by effective gas extraction
    • Simple calibration
    • Software: simple, clear, intuitive
    • Definition of own methods for device control
    • Typical measuring time 5 min (depending on sample properties)

    Applications

    • Water, drinking water, surface water
    • Municipal wastewater
    • Industrial wastewater
    • Monitoring of landfill-leachate
    • Gas analysis (e. g. LNG, LPG)
    • H2S in hydrocarbon mixtures
    • Investigation of technical and pharmaceutical products (e. g. storage stability)
    • Quality management

    Advantages

    • Analysis of the original sample
    • No sample preparation
    • Dosing manually or optional fully automatic
    • Minimized cross sensitivity through the indirect method
  • Measuring range 0.01 ... 10,000 ppm (dependent on sample volume)
    Resolution 0.1 μg abs., output signal linear
    Measuring duration 2 ... 10 min (dependent on the sample), usually 5 min
    Sample volume 0.01 ... 20 mL
    Gas flow Up to 50 L/h
    Power supply 230 V/50 Hz, 115 V/60 Hz
    Power input 150 W
    Dimensions 480 x 390 x 290 mm (W x D x H)
    Weight 11 kg
    Data connection RS 232 / USB (with converter)
    Device control PC software (PC not included in the scope of delivery)