THERMOSTATIC CALIBRATORS 型號:PYROS 650
  • Temperature range: 35 °C ÷ 650 °C
  • Precision: ±0,50 °C
  • Hole depth: 150 mm
  • Hole diameter: 26 mm
  • Operating range:-24 °C ÷ +650 °C, (-11 °F ÷ 1202 °F)

    Applications:

    Control and calibration of temperature sensor, in the laboratory and in the field, in conformity with ISO 9000 standards
    Control of thermostats.
    Automatic computer-controlled calibrations.

  • Technical specification PYROS 140 PYROS 375 PYROS 650
    Operating range
    (t amb. 20 °C / 68 °F)
    -24 °C ÷ 140 °C
    (-11 °F ÷ 284 °F)
    amb.+10 °C ÷ +375 °C (Amb. +18 °F ÷ 707 °F) amb.+15 °C ÷ +650 °C (Amb. +27 °F ÷ 1202 °F)
    Mean heating time (stabilization included) From amb. to 120 °C / 248 °F 20 min From 30 °C / 86 °F to 375 °C / 707 °F
    20 min
    From 50 °C / 122 °F to 650 °C / 1202 °F
    35 min
    Mean cooling time (stabilization included) From amb. to -20 °C / -4 °F 17 min From 375 °C / 707 °F to 100 °C / 212 °F
    40 min
    From 650 °C / 1202 °F to 100 °C / 212 °F
    60 min
    Axial uniformity ±0,05 °C at –20 °C
    ±0.09 °F at –4 °F
    ±0,10 °C at 50 °C
    ±0.1 °F at 122 °F
    ±0,02 °C at 50 °C
    ±0.03 °F at 122 °F
    ±0,13 °C at 250 °C
    ±0.2 °F at 482 °F
    ±0,04 °C at 0 °C
    ±0.07 °F at 32 °F
    ±0,20 °C at 150 °C
    (±0.3 °F at 302 °F)
    ±0,05 °C at 150 °C
    (±32.09 °F at 302 °F)
    ±0,15 °C at 450 °C
    (±32.27 °F at 842 °F)
    ±0,10 °C at 100 °C
    (±32.1 °F at 212 °F)
    ±0,30 °C at 375 °C
    (±32.5 °F at 707 °F)
    ±0,15 °C at 375 °C
    (±0.27 °F at 707 °F)
    ±0,35 °C at 650 °C
    (±0.6 °F at 1202 °F)
    Radial uniformity (at 40 mm) Not available ±0,10 °C at 50 °C
    (±0.1 °F at 122 °F)
    ±0,05 °C at 50 °C
    (±0.09 °F at 122 °F)
    ±0,22 °C
    (±0.39 °F)
    ±0,15 °C at 150 °C
    (±0.27 °F at 302 °F)
    ±0,10 °C at 150 °C
    (±0.1 °F at 302 °F)
    ±0,20 °C at 375 °C
    (±0.3 °F at 707 °F)
    ±0,15 °C at 375 °C
    (±0.27 °F at 707 °F)
    Hole depth 104 mm (4 in.) 150 mm (5.9 in.) 150 mm (5.9 in.)
    Hole diameter 2 x 13 mm (0.07 x 0.51 in)
    1 x 19 mm (0.03 x 0.74 in)
    26 mm (1 in.) 26 mm (1 in.)
    Display accuracy (±1 digit)   ±0,25 °C at 100°C
    ± 0.45 °F at 212 °F
    ±0,25 °C at 150 °C
    (±0.45 °F at 302 °F)
    ±0,25 °C at 150 °C
    (±0.45 °F at 302 °F)
    ±0,50 °C
    (±0.9 °F)
    ±0,35 °C at 375 °C
    (±0.6 °F at 707 °F)
    ±0,35 °C at 375 °C
    (±0.6 °F at 707 °F)
    Stability ±0,05 °C - (± 0.09 °F) ±0,15 °C - (±0.27 °F) ±0,10 °C - (±0.1 °F) ±0,30 °C - (±0.5 °F)
    Display Resolution 0,1 °C - (0.1 °F) 0,1 °C - (0.1 °F) 0,1 °C - (0.1 °F)
    Units of measure °C / °F °C / °F °C / °F
    Over temperature cut out No Yes Yes
    Switch test 12 VAC DC 5 VAC DC 5 VAC DC
    PC Interface RS 232 RS 232 RS 232
    Display LED double line LED double line LED double line
    IN / OUT RS232 RS232 RS232
    Temperature ramps Yes Yes Yes
    Calibration From 1 to 10 calibration points From 1 to 10 calibration points From 1 to 10 calibration points
    Calibrator size 130 x 260 x 280 mm
    (5.1 x 10 x 11 in.)
    130 x 260 x 280 mm
    (5.1 x 10 x 11 in.)
    130 x 260 x 280 mm
    (5.1 x 10 x 11 in.)
    Calibrator weight 4,9 kg - (10 lb) 5,4 kg - (11 lb) 6,0 kg - (13 lb)
    Power supply Voltage supply 100/230 VAC ±10% - 50/60 Hz Automatic switchable voltage supply 115/230 VAC ±10% - 50/60 Hz Automatic switchable voltage supply 115/230 VAC ±10% - 50/60 Hz
    Electric power 80 W 600 W 600 W
    Diameter of the sensor used: PYROS140-1H: Pt100 probe Ø 3 mm                                   
    PYROS375: Pt100 probe Ø 6 mm / Pt100 probe Ø 3 mm
    PYROS140-2H: Pt100 probe Ø 3 mm                                   
    PYROS650: TcN probe Ø 4,5 mm