0 like search mail home
選擇產品功能 arrow_down
>

PAT-Cell-Press

PAT-Cell-Press 型號:PAT-Cell-Press

Leakproof test cell for measuring gas evolution.

  • PAT-Core design with or without reference electrode
  • Laser welded pressure sensor, 0 to 3 bar abs.
  • Electrode feedthroughs with glass-to-metal seals
  • The PAT-Cell-Press is the ideal, leakproof cell for pressure testing.

    Our pressure test cell for the PAT series
    The PAT-Cell-Press is a pressure test cell for measuring gas evolution and drawing gas samples (option). Its advanced design includes a laser welded pressure sensor and glass-to-metal seals. The number of non-permanent seals is reduced to its minimum: a single seal between cell lid and base and a ferrule when the cell is ordered with the optional gas sample port.

    In order to ensure best testing results, every PAT-Cell-Press is intensively tested for helium leaks.

    The PAT-Cell-Press can be operated directly with the PAT-Tester-x and PAT-Tester-i-16 potentiostats. If used with a 3rd party potentiostat, the PAT-Press-Box is required.

    Sample test results

  • Diameter (Lid) 49.5 mm
    Diameter (Cell base) 44 mm
    Height 73 mm
    Width 70 mm (with gas sample port)
    Weight 0.5 kg
    Separator diameter 21.6 mm
    Electrode diameter 18 mm
    Operational temperature -20 to +80°C
    Pressure sensor:  
    Range 0 to 3 bar abs. (0.5 to 4.5 V output signal)
    Accuracy < 5 mbar
    Resolution < 0.2 mbar


    Dead volume (PAT-Cell-Press with standard lid)

     
    Test cell without PAT-Core 8.144 ml
    Test cell with inserted PAT-Core 3.565 ml