ECD-4-nano
ECD-4-nano 型號:ECD-4-nanoAdvanced Electrochemical Dilatometer for the measurement of electrode expansion with nanometer resolution.
- Capacitive displacement sensor (range 250 μm, resolution better than 5 nm)
- Additional gas pressure (0 to 3 bar) and temperature sensor (-20 to 80° C)
- Cableless connection via PAT socket, with electronic cell tag (PAT-Button)
- PRODUCT DETAILS
-
Advanced test cell for the measurement of the electrode expansion in the nanometer range.
The ECD-4-nano takes dilatometry measurements to the next level
The ECD-4-nano is a high-resolution electrochemical dilatometer. The most powerful version offers a capacitive parallel plate sensor system with a resolution of better than 5 nanometers. This makes the ECD-4-nano the perfect instrument for detecting thickness changes of the individual electrode or the full cell stack during the electrochemical cycle.
The ECD-4-nano‘s completely redesigned test cell features a corrosion-resistant cell bottom and a new One-Seal concept that significantly improves tightness over previous ECD-3 models. This enables stable long-term operation as well as the use of a wide range of electrolytes.
To further improve workflow and handling, we have now integrated the ECD-4-nano into the PAT system. The dilatometer can be inserted directly into a PAT-Tester-x or a docking station such as the PAT-Clamp-1. This allows a space-saving and fast setup of the instrument. Needless to say, the integrated PAT-Button also ensures automatic recognition of the test cell in our EL-Software measurement software.
Sample Test Results
Sample Setup
Sample SetupWorking principle of the ECD-4-nano (3-electrode setup)
he working electrode (WE) and the counter electrode (CE) are separated by a stiff glass frit soaked with electrolyte. The upper WE is sealed by means of a flexible metal membrane, through which any charge-induced thickness change is transmitted towards the sensor/load unit attached on top. The fixation of the glass frit ensures that only the thickness change of the working electrode is being detected without interference from the CE. -
Length 64 mm Width 67 mm Height 149 mm Weight 2 kg Sample diameter 10 mm Glass T-Frit (Separator) 12.5/10 x 3.5 mm Separator diameter (full cell mode) max. 10 mm Working (upper) electrode diameter max. 10 mm Counter (lower) electrode diameter max. 12 mm Gas pressure sensor: 0 to 3 bar abs. Chemical compatibility Aprotic organic electrolytes Cell electrolyte volume approx. 0.2 ml Operational temperature range (cell and sensor) -20 to 80° C Operational temperature range (conditioning electronics and data logger) 0 to 40 °C Load on test specimen approx. 1 Newton Product variants ECD-4-nano ECD-4 Displacement sensor type: Capacitive LVDT Displacement range: 250 μm 500 μm Displacement resolution: ≤ 5 nm ≤ 50 nm Compatible PAT docking stations and potentiostats