多功能校正器
Multifunction calibrator 型號:C101Series of small dimensions calibrators of AC and DC voltage and current up to 1100V and 100A, with high load possibility.
It allows checking of economical multimeters and clamps meter
-
- Voltage source up to 1100V
- Current source up to 20A
- Small dimensions and light weight - 16kg
- High output power
- Option C101FC with 100A AC subrange
- Output power 40VA for 20A AC/DC
- Software Calpro 101
- Interface RS232C
- Calibration Certificate
The C101F Calibrator enables generating of:
- direct (DC) and alternating (AC) voltages up to 1100V in five subranges 0.1-1-10-100- 1000V,
- DC and AC currents up to 20.5A in five subranges 0.001-0.01-0.1-1-10A (the C101FC option up to 100A AC).
Frequency of alternating signals may be chosen from values 50Hz (synchronized with power supply frequency), 60Hz, 400Hz or can be programmed in range 45.00...2000Hz. Output signal can be set by means of external keyboard or by standard interface RS232C from PC computer.
The C101F Calibrator is characterised by high accuracy, high output power, small dimensions and light weight. -
Technical parameters of C101F Calibrator
Parameter
Range
100mV
1V
10V
100V
1000V
1mA
10mA
100mA
1A
10A
Resolution
100nV
1mV
10mV
100mV
1mV
1nA
10nA
100nA
1mA
10mA
Settings range
1...
205mV0.01...
2.05V0.1...
20.5V1...
205V10...
1100V0.01...
2.05mA0.1...
20.5mA1...
205mA0.01...
1.1A0.1...
20.5AUncertainty *) DC [% of set value+
0.03
+0.0050.02
+0.0020.02
+0.0030.03
+0.003% setting range]
AC
0.05
+0.005+50mV0.05
+0.005+0.5mA0.05
+0.005Maximum load
20mA
1100mA
200mA
30mA
11V
2V@20A
4V@10AFrequency
DC; 50Hz synchronized with power supply; 60±0.01Hz and 400±0.1Hz 45.00...99.99±0.01Hz, 100.0...999.9±0,1Hz, 1000...2000±1Hz
Temperature coefficient
±0.1 of uncertainty value / 1°C
in environment temperature +5...+18°C and +28...+40°CWeight and dimensions
16kg and 478 (width)x194 (height)x342 (depth) mm
Power supply
230V ±10% / 45…65Hz / 200VA
DC – direct voltage and current, AC – alternating voltage and current
*) Absolute Uncertainty with 95% Confidence Level includes the traceability to external standards,12th months stability, influences of ambient temperature, power supply voltage and load regulation in reference conditions and nonlinearity. For frequency band above 500Hz-linear rise up to double value for frequency 2kHz, typically.