Water extraction of the samples at temperatures 35 °C up to 250 °C, e. g.:
- Constant temperature
- Individually set up temperature programs
- Freely selectable temperature ramp
- Time-controlled temperature programs for step-by-step heating
- Type of result: µg, µg/L, mg/L, mg/kg, ppm, %, by using the formula generator
Areas of Application
- Transformer oils
- Hydraulic oils
- Engine oils
- Gear oils
- Crude oil
- Fuel oil
- Gasoline
- Diesel
- Jet fuel
- Bio fuel
- Petroleum products
- Insulating oils
- Silicone oil
- Lubricating oils
- Biopetroleum
- Biological oils
Advantages
- Closed loop principle does not allow methanol to evaporate from the KF solvent
- Reagent capacity is used completely
- Additive and Sulphur side reactions minimized
- By using of temperature programs it is possible to separate free and chemically bonded water
- Temperature ramping program allows you to distinguish between various types of bonded water
- No blank value meaning true ppm accuracy
- Aquamax KF PRO Oil can be used in the laboratory or used as part of a mobile lab when taking a measurement from the sample point is critical
- Compact and rugged device
-
Measurement method: Coulometric Karl Fischer titration Sample administration: Manually with syringe Sample amount: 0.01 - 20 mL Heating temperature: 35 ... 250 °C, isothermal or with temperature program Blank value: 0 µg Measuring range: 0.0001 ... 100 % Power supply: 230 V/50 Hz; 115 V/60 Hz Dimensions: 33 x 39 x 48 cm (W x D x H) Weight: 17.0 kg