Sulfimax GX online WATER 型號:Sulfimax GX online WATER

Analyzer for online determination of sulphides/H2S in wastewater even below the odor threshold

  • Online sulphide determination in water and wastewater

    Dosing chemicals as required.
    Save costs and protect the environment.

    The Sulfimax GX online WATER continuously measures hydrogen sulphide (H2S) and sulphides in wastewater and other other aqueous samples quickly and accurately. This is important for timely and on-demand intervention in process streams and wastewater treatment procedures.

    Effective gas extraction completely expels H2S from the sample. Interference from the sample matrix is virtually non-existent. The released H2S gas is directed to the highly sensitive sensor, which detects H2S in the range of 0.01 to 1000 ppm.

    A measurement takes 5 to 15 min, depending on the sample composition.

    With the Sulfimax GX online WATER, even industrial wastewater that is basic or contaminated with hydrocarbons or ammonia can be monitored reproducibly.

    Sample aspiration, all rinsing steps and the return to the pipe system are performed automatically. Configurations can be edited and results read out via remote access. The measured values can be be transmitted to on-site alarm systems via digital and analog outputs.

    The analyzer can be used as a H2S dependent controlling system for chemical wastewater treatment.

    The Sulfimax GX online WATER as ATEX compliant version is suitable for use in ATEX zone 1 (gas) according to DIN EN IEC 60079-2.

    Features

    • Detection of the true H2S content directly in the liquid phase and thus independent of weather conditions (even in strongly basic samples)
    • Simple calibration
    • Simple, clear software
    • No cross-contamination due to self-cleaning between measurements
    • Long sensor life due to integrated regeneration processes

    Applications

    • Online aquisition of the current sulphide concentration
    • Environmental analytics, e. g.:
      - Waste water analysis for control systems
      (pump sump, free level or pressure water pipe)
      - Landfill leachate monitoring
      - Industrial sewage treatment plants
      - Municipal sewage treatment plants

    Advantages

    • Automatic sample dosing
    • Hardly any cross-sensitivities due to indirect method
    • Output 4 - 20 mA for integration of the H2S signal into the control system
    • Short measuring intervals allow fast reaction to changes
    • Low maintenance requirements
  • Sample dosing Via transfer line (up to 30 m long, from up to 5 m depth)
    Typical duration 5 ... 15 min (depending on the sample)
    Range 0.01 ... 1000 ppm
    Resolution 0.1 µg abs., output signal linear
    Sample volume 0.01 ... 20 mL
    Gas  supply Internal pump or compressed air
    Gas  flow Up to 50 L/h
    Power supply 220 ... 230 V/50 Hz, 2 A
    Power consumption 250 W
    Type of protection IP 54
    Dimensions 600 x 370 x 720 mm (W x D x H)
    Weight 30 kg