0 like search mail home
選擇品牌 arrow_down

PFP7 Industrial Flame Photometer

PFP7 Industrial Flame Photometer 型號:PFP7
• Designed for industrial analysis
• Supplied with Na, K, Li, Ba and Ca filters
• Low temperature, single channel
• Flame failure safety system
• Operates with propane, butane, natural gas or LPG
  • The PFP7 is a low temperature, single channel flame photometer that is designed for the routine determination of sodium, potassium, calcium, barium and lithium concentrations. The flame failure safety system makes these products ideal for use in industrial and educational environments. The use of fine and coarse sensitivity controls allows for accurate measurements each and every time.
  • Parameter

    PFP7

    Range

    0 to 199.9ppm

    Limits of detection Na

    0.2ppm

    K

    0.2ppm

    Li

    0.25ppm

    Ca

    15ppm

    Ba

    30ppm

    Reproducibility

    <1% coefficient of variation for 20 consecutive samples using 10ppm Na set to read 50

    Linearity

    < 2% error when 3ppm Na/K and 5ppm Li are set to read 100

    Stability

    < 2% drift over 5min when continuously aspirating 10ppm sample set to 50.0. Specificity Interference from Na/K and Li equal in concentration to test element will be <0.5%

    Recorder output

    Nominal 1.00V for a reading of 100.0

    Electrical supply

    90-125V or 190-250V @50/60Hz

    Air supply

    Moisture and oil free, 6 litre/min @ 14psi

    Fuel

    Propane, butane, natural gas or LPG

    Size (w x d x h)

    420 x 360 x 300mm

    Weight

    8kg