- Neware
- Pine Research
- OKOndt Group
- Novocontrol
- Norecs
- METERTEST
- Microrad
- Metrel
- Labdex
- SDT
- micrux
- AARONIA AG
- ADASH
- Amptek
- AOiP
- AstroNova
- AWSensors
- Automatic Research
- BASI
- BRS
- Cmc
- CTRL
- CALMET
- CHECKLINE
- C-Tech
- DV Power
- DANATRONICS
- ECH
- Elsys
- Enervac
- Enapter
- ELVEFLOW
- EA Technology
- EL-CELL
- ENERGY SUPPORT
- Electrothermal
- FASTEC
- GE
- GMW
- Gaskatel
- GIUSSANI
- Globecore
- GREENLIGHT
- GRZ
- HTW
- HIGH SENSE SOLUTIONSHTW
- HUBER
- Labnics
- Ida
- LIQUID
- Instytut Fotonowy
- KEHUA TECH
- JGG
- HVPD
- Jenway
- Jacomex
- IVIUM
- ndb
- OZM
- Redoxme
- Serstech
- SATIR
- VacCoat
- Zurich
- Neware
- Pine Research
- OKOndt Group
- Novocontrol
- Norecs
- METERTEST
- Microrad
- Metrel
- Labdex
- SDT
- micrux
- AARONIA AG
- ADASH
- Amptek
- AOiP
- AstroNova
- AWSensors
- Automatic Research
- BASI
- BRS
- Cmc
- CTRL
- CALMET
- CHECKLINE
- C-Tech
- DV Power
- DANATRONICS
- ECH
- Elsys
- Enervac
- Enapter
- ELVEFLOW
- EA Technology
- EL-CELL
- ENERGY SUPPORT
- Electrothermal
- FASTEC
- GE
- GMW
- Gaskatel
- GIUSSANI
- Globecore
- GREENLIGHT
- GRZ
- HTW
- HIGH SENSE SOLUTIONSHTW
- HUBER
- Labnics
- Ida
- LIQUID
- Instytut Fotonowy
- KEHUA TECH
- JGG
- HVPD
- Jenway
- Jacomex
- IVIUM
- ndb
- OZM
- Redoxme
- Serstech
- SATIR
- VacCoat
- Zurich
高級記錄多功能溫度校正器
Advanced documenting multifunction calibrator thermometer 型號:CALYS 150-
CALYS 150現場記錄多功能校正器是該系列的頂級儀器。 它是在所有行業中進行高級過程維護和在測試台上使用的完美工具。 它適用於所有現場和實驗室測量,可在一台儀器中同時測量和產生兩個隔離通道的溫度,電阻,製程,壓力和頻率等各種訊號。 CALYS 150不僅可用作模擬器(IN / OUT),還可用作雙通道溫度計(IN / IN)進行比較校正。 它校準HART變送器(整合在ACL500調變解調器中的HART通訊器)和熱敏電阻。 提供擴展功能(溫度模擬,縮放,步驟,合成器,統計功能...)和稽核追踪,CALYS 150符合21 CFR Part 11和NADCAP熱處理標準,使高級數據利用和完整數據可追溯性更容易。 CALYS 150具有高性能,適用於高級用途:
- 溫度高達0.005%RDG。
- 電阻高達0.006%RDG和50KΩ範圍。
- 電流:RDG和100 mA範圍高達0.007%,+ 24 V迴路電源。
- 電壓:RDG高達0.005%,50 V範圍。
- 頻率:RDG和100 KHz範圍高達0.01%。
- 壓力:外部壓力模組(與壓力泵進行比較校正)。
使用這種用戶友好型儀器,可以在整個過程鏈中快速執行校正任務。 在整週內將900 g記錄過程的校正器帶到現場,並在設備中儲存10個校正程序。 將探頭連接到儀器(Easyconnectystem®)後執行程序並儲存結果,以便現場輕鬆快速校正。 回到辦公室,然後您可以在電腦上傳數據,以便發布自訂校準證書使用專用校準軟體DATACAL。 IP 54由防電磁干擾的橡膠套完全保護,CALYS 150整合了“easyconnect”端子和寬背光顯示屏,便於在任何惡劣或黑暗條件下使用。 當與外部壓力模組(參考ACL433)一起使用時,CALYS 150可以測量和模擬壓力(使用壓力幫浦進行比較校正)。 當與特定電纜相關時,CALYS 150也具有驅動浴槽和乾燥塊的能力(參考ACL600)。
易於連接的系統
只需按下端子頂部並插入最大直徑為3 mm或10 AWG的電線和補償熱電偶連接器即可連接探頭。 導線緊固在兩塊黃銅板之間,確保熱穩定性,並為熱電偶提供非常好的冷端補償。 該系統還可以在終端頂部連接4 mm香蕉插頭和安全連接器。
連接感測器時,只需按下端子頂部,然後插入長達 3 mm或 10 mm的導線即可。
CALYS系列,4種型號,從基本使用到高級性能 -
溫度@ 23°C±5°C時的性能和技術規格
不確定度以讀數百分比(CALYS 150顯示)+固定值給出。
電阻探針:測量和類比2線,3線或4線電阻探頭測量:自動識別連接線數,螢幕上顯示4線安裝探頭的準確度考慮所用溫度感測器的特殊誤差和實施條件允許的測量電流:0.01 mA至4 mA在類比模式下,給出1 mA測量電流(Pt50 / 100,Ni100 / 120,Cu10 / 50)或0.1 mA(Pt200 / 500/1000,Ni1000)的規格溫度係數:<10%精度/°C
熱電偶:測量和類比熱電偶:PlatineL,Mo,NiMo / NiCo,G,D,L,C:有關規格,請參閱使用說明書(可根據要求提供)精度參考@ 0°C。 使用內部參考結(B雙除外)時,在0°C時額外增加0.2°C的不確定度。
可以通過程式設計冷端定位來選擇(熱電偶B除外):外部溫度為0°C,內部(儀器端子的溫度補償)或手動輸入。
溫度係數:<10%精度/°C顯示單位:°C和F.熱敏電阻:測量(通道1)
通過集成50 Kohms系列和Steinhart - Hart方程,可將熱敏電阻輸入CALYS 150並進行測試。 Steinhart-hart方程式如下:1 = A + B(ln(R))+ C(ln(R))3 T其中:A,B和C通常根據溫度在0°C,25°C下計算和70°C壓力規格和性能@ 23°C±5°C
壓力:通過外部數位感測器測量提供相對壓力、絕對壓力和差壓。 介面:¼ 氣壓解析度:0.02% FS 0.02% FS 精確度:-0.05% FS(10 至40°C)-0.1% FS(-10 至+10°C,40 至80°C) 此數位壓力模組ACL433 透過RS485 序列電纜連接至CALYS 150 的數位輸入連接器。 所有數據均為數字數據。 測量值在出廠時透過韌體中的多項式修正進行溫度補償。
工作溫度為 23°C ±5°C 時的規格和性能
直流電流:測量
有無迴路供電溫度係數:<7 ppm /°C,0°C至18°C和28°C至50°C迴路電源:24 V±10%HART®相容性:輸入阻抗Rin =280Ω帶線性或二次刻度的顯示
直流電壓:測量
頻率,計數:測量
刻度單位:脈衝/分鐘和Hz觸發電位:1 V測量頻率信號或乾接點。 計數將在規定的時間或無限時間內進行。
電阻:測量
2線,3線或4線電阻測量:自動識別連接線數,螢幕上顯示4線安裝探頭的精確度。
直流電流:發射
有無迴路供電
溫度係數<7 ppm /°C,0°C至18°C和28°C至50°C CALYS 150配置規格如下:
活動模式(+ 24V ON)1被動模式(+24 V OFF) )
被動模式(+24 V OFF)1儀錶處於活動模式(+24 V ON)
預程式設計步驟 :
0%25%50%75%100%
4-20 mA線性4 8 12 16 20
0-20 mA線性0 5 10 15 20
4-20 mA quad 4 5 8 13 20
0-20 mA quad 0 1.25 5 11,25
20 4-20 mA valves 3.8-4-4.2 12 19,20,21直流電壓:發射
頻率,脈衝:發射
刻度單位: 脈衝/分鐘和Hz 脈衝發射和乾接點模擬。 最大振幅:20 V(使用者可選)。
電阻:發射
脈衝電流發射:規格請參閱使用說明書 溫度係數: < 5 ppm/°C(0°C 至 18°C,28°C 至 50°C)。 電流建立時間: <1ms 與智慧變送器相容 Iext:校正器接收的電流。
-
CALYS 150 field documenting multifunction calibrator is the top instrument of the range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument. CALYS 150 does not only work as a simulator (IN / OUT) but also as a dual channel thermometer (IN / IN) to perform comparison calibration. It calibrates HART transmitters (HART communicator integrated into ACL500 modem) and thermistors. Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, CALYS 150 complies with both 21 CFR Part 11 and NADCAP Heat Treatment standards and makes advanced data exploitation and full data traceability easier. High performances for CALYS 150, for advanced use:
- Temperature Up to 0.005 % RDG
- Resistance Up to 0.006 % RDG and 50 KΩ range
- Current: Up to 0.007 % RDG and 100 mA range + Loop Supply 24 V
- Voltage: Up to 0.005 % RDG and 50 V range
- Frequency: Up to 0.01 % RDG and 100 KHz range
- Pressure: With an external pressure module (comparison calibration with a pressure pump)
Using this user-friendly instrument, calibration tasks can be quickly carried out over the whole process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument (Easy connect system®) and save the results for onsite easy and quick calibration. Back to the office, you can then upload the data on a computer in order to issue customized calibration certificates with dedicated calibration software DATACAL. IP 54, fully protected by an antichoc rubber holster, CALYS 150 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions. When used with an external pressure module (ref. ACL433), CALYS 150 can measure and simulate pressure (comparison calibration with a pressure pump). CALYS 150 has also the capability to drive baths and dry-blocks when associated with the specific cable (ref. ACL600).
Easy connection system
Connect your probes by simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors. Wires are held tight between two brass plates ensuring thermal stability and a very good cold junction compensation for thermocouples. This system also enables 4 mm banana plugs and security connectors to be connected on the terminal top.
CALYS series, 4 models from basic use to advanced performances
-
Specifications and performances in temperature @23°C ±5°C
Uncertainty is given in % of reading (CALYS 150 display) + fixed value.Resistive probes: Measurement and simulation
Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen Accuracies are given for 4-wire mounted probes Take into account particular error of temperature sensor used and implementation conditions Admissible measuring current: 0.01 mA to 4 mA In simulation mode, specifications given for 1 mA measuring current (Pt50 / 100, Ni100 / 120, Cu10 / 50) or 0.1 mA (Pt200 / 500 / 1000, Ni1000) Temperature coefficient: < 10% of accuracy /°C
Thermocouples: Measurement and simulation
Thermocouples: PlatineL, Mo, NiMo/NiCo, G, D, L, C: For specifications, refer to the instruction manual (Available on request) Accuracy is given for reference @ 0°C. When using the internal reference junction (except couple B) add an additional uncertainty of 0.2 °C at 0 °C. It is possible (thermocouple B excepted) to choose by programming the cold junction localization: External at 0°C, internal (temperature compensation of instrument’s terminals) or manually entered. Temperature coefficient: <10% of accuracy /°C Display unit: °C and F.
Thermistors: Measurement (Channel 1)With 50 Kohms range and Steinhart – Hart equation integrated, thermistors can be entered into CALYS 150 and tested. Steinhart-hart equation is as follows: 1 = A + B (ln(R)) + C(ln(R))3 T Where: A, B and C are usually calculated according to temperature at 0°C, 25°C and 70°C
Specifications and performances in pressure @23°C ±5°C
Pressure: Measurement by external digital sensor
Available in relative, absolute and differential pressure. Connector: ¼ gas Resolution: 0.02% FS Accuracy: -0.05% FS from 10 to 40°C - 0.1% FS from -10 to +10°C and from 40 to 80°C This digital pressure module ACL433 is connected to CALYS 150 through RS485 serial cable to the digital input connector. All data are digital. Measurements are compensated in temperature by a polynomial correction implemented into the firmware at factory.Specifications and performances in process @23°C ±5°C
DC current: Measurement
With or without loop supply
Temperature coefficient: < 7 ppm/°C from 0°C to 18°C and 28°C to 50 °C Loop supply: 24 V ± 10% HART® compatibility: Input impedance Rin = 280 Ω Display with linear or quadratic scalingDC voltage: Measurement
Frequency, counting: Measurement
Scale unit: Pulse / min and Hz Trigger level: 1 V Measurement on frequency signals or dry contacts. Counting will be performed on defined time or infinite time.
Resistance: Measurement
Resistance measurement in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen Accuracies are given for 4-wire mounted probes
DC current: Emission
With or without loop supplyTemperature Coefficient < 7 ppm/°C from 0°C to 18°C and 28°C to 50 °C Specifications given for CALYS 150 configurations in:
Active mode (+24V ON) 1 Meter in passive mode (+24 V OFF)
Passive mode (+24 V OFF) 1 Meter in active mode (+24 V ON)
Pre-programmed steps 0% 25% 50% 75% 100%
4-20 mA linear 4 8 12 16 20
0-20 mA linear 0 5 10 15 20
4-20 mA quad 4 5 8 13 20
0-20 mA quad 0 1.25 5 11,25 20
4-20 mA valves 3.8-4—4.2 12 19, 20, 21DC voltage: Emission
Frequency, pulse: Emission
Scale unit: Pulse / min and Hz Pulse emission and dry contacts simulation. Max. amplitude: 20 V (User selectable)
Resistance: Emission
Emission with pulsed current available: refer to the instruction manual for specifications Temperature coefficient: < 5 ppm/°C from 0°C to 18°C and 28°C to 50 °C. Current establishing time: <1ms Compatibility with smart transmitters Iext : Current received by the calibrator